Features
1Design patent,software patent
2 Application of CSE database,reliable measurement and analysis results
3 12-lead simultaneous acquisition
4 6"LCD,real-time display of 12-lead EGC waveforms
5 Detection and alarm on lead-off and low battery
6 Automatic adjustment of baseline for optimal printing
7Built-in EGC simulator for DEMO purpose
8 3kinds of operation mode:AUTO,MAN and ANA
9 3kinds of filter :HUM,EMG and ADS
10 AC,DC and battery power supply
11 2 types lead mode:Standard and Cabrera
12 1,000 EGC files can be saved to SD card
13 Communication with PC(Option)
Specification
Safety Standard:IEC Class 1,type CF
Acquisition Mode:12 lead acquisition simultaneously
Input Circuit:Floating,protection against defibrillator and pacemaker
A/D converter:12 bit
Time Constant:≥ 3.2 s
Frequency Response:0.05Hz~165Hz(-3dB)
Sensitivity: 2.5,5,10,20,40(mm/mV)Auto
Input Impedance:≥50MΩ
Input Circuit Current:≤50nA
Calibration Voltage:1mV±2%
Noise Level:<15µVp-p
Anti Baseline Drift :Automatic
Filter:EMG filter 35Hz~45Hz(-3dB)
ADS filter:Auto( 0.15~0.50Hz)
HUM filter:50Hz/60 Hz(-20dB)
CMRR:>100dB
Patient Current leakage:<10µA
Paper Speed:6.25/12.5/25/50 mm/s
Recording Paper:112mm(width)20/30mm(length),roll
Power Supply:AC:110/220V 50/60Hz;85V-265V (Option)
DC:14.4V ,built-in rechargeable lithium batter
Dimension :345mm*300mm*80mm;415mm*195mm*380mm(Carton)
Net Weight:2.5kgs